An Automatic Calibration Method for Near-infrared Camera in Optical Surgical Navigation

RongQian Yang, Xuan Si, QinYong Lin, Ken Cai

Abstract


Optical surgical navigation system (SNS) with near-infrared tracking system is becoming extensively used in clinics, and the accuracy of SNS is influenced by the calibration of near-infrared cameras (NIRCs). We propose an automatic calibration method for NIRCs. The method is based on a designed calibration board. In our experiments, corners are automatically extracted to obtain the parameters of NIRCs. This method has the advantages of saving time, efficiency in computation, high accuracy, and reliability. In our experiments, an NIRC can be calibrated in only 5 s. Meanwhile, the average relative errors of the focal length and principal point are 0.87% and 1.39%, respectively.

Full Text:

PDF


DOI: http://dx.doi.org/10.12928/telkomnika.v13i4.3103

Article Metrics

Abstract view : 134 times
PDF - 155 times

Refbacks

  • There are currently no refbacks.


Copyright (c) 2015 Universitas Ahmad Dahlan

TELKOMNIKA Telecommunication, Computing, Electronics and Control
ISSN: 1693-6930, e-ISSN: 2302-9293
Universitas Ahmad Dahlan, 4th Campus, 9th Floor, LPPI Room
Jl. Ringroad Selatan, Kragilan, Tamanan, Banguntapan, Bantul, Yogyakarta, Indonesia 55191
Phone: +62 (274) 563515, 511830, 379418, 371120 ext. 4902, Fax: +62 274 564604

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

View TELKOMNIKA Stats